รายละเอียดสินค้า 572-2
High Temperature Infrared Thermometer
|
-
Measure -30 °C to 900 °C (-22 °F to 1652 °F)
-
60:1 distance to spot ratio with dual laser sighting for fast, accurate targeting
-
Current Temperature plus MAX, MIN, DIF, AVG temperature displays
-
Compatible with all standard mini-connector K-type thermocouples, including ones you already own and have installed.
-
High and low temperature alarm
-
USB 2.0 computer interface cable
-
FlukeView Forms Documenting Software
-
Two-year warranty
Specification |
IR Temperature Range |
-30 °C to 900 °C (-22 °F to 1652 °F) |
IR Accuracy
Calibration geometry with ambient temperature 23 °C±2 °C) |
<0 °C: ±(1.0 °C + 0.1 °/1 °C)
≥0 °C: ±1 % of reading or ± 1.0 °C, whichever is greater
<32 °F: ±(2 °F ±0.1 °/1 °F)
≥32 °F: ±1 % or ±2 °F, whichever is greater |
IR Repeatability |
±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater. |
Display Resolution |
0.1 °C / 0.1 °F |
Spectral Response |
8 μm to 14 μm |
Response Time (95 %) |
<500 ms |
K-Type Thermocouple Input Temperature Range |
-270 °C to 1372 °C (-454 °F to 2501 °F) |
K-Type Thermocouple Input Accuracy
(with ambient temperature 23 °C±2 °C) |
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is |
K-Type Thermocouple Repeatability |
±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater |
K-Type Thermocouple Input Accuracy (with ambient temperature 23 °C±2 °C) |
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is greater
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is |
K-Type Thermocouple Repeatability |
±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater |
K-Type Thermocouple Resolution |
0.1 °C/0.1 °F |
Distance:Spot |
0:1(calculated at 90 % energy) |
Laser sighting |
Offset dual laser, output <1 mW |
Data storage |
99 points |
Power |
2 AA Batteries
|
|