รายละเอียดสินค้า Fluke 725Ex Intrinsically Safe Multi-function Process Calibrator
|
Measurement Accuracy |
|
Voltage DC |
|
30.000 V |
0.02%+ 2 counts (upper display) |
|
30.000 V |
0.02%+ 2 counts (lower display) |
|
100.00 mV |
0.02%+ 2 counts |
|
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
|
|
Current DC |
|
24.000 mA |
0.02%+ 2 counts |
|
|
Resistance |
|
0.0 to 400.0 Ω |
0.1Ω (4-wire), 0.15Ω (2- and 3-wire) |
|
401 to 1500 Ω |
0.5Ω (4-wire), 1Ω (2- and 3-wire) |
|
1500 to 3200 Ω |
1Ω (4-wire), 1.5Ω (2- and 3-wire) |
|
|
Frequency |
|
2.0 to 1000.0 CPM |
0.05% |
|
1.0 to 1100.0 Hz |
0.05 % + 1 count |
|
1.00 to 10.00 kHz |
0.05 % + 1 count |
|
Sensitivity |
1 V peak-to-peak-minimum |
|
|
Pressure |
|
Accuracy |
from 0.025% of range using any of 29 pressure modules. (for detailed specifications refer to pressure modules in options and accessories) Modules available for differential, gage, vacuum, absolute, dual and high pressure. |
|
|
Source Accuracy |
|
Voltage DC |
|
100.00 mV |
0.02%+2 counts |
|
10.000 V |
0.02%+2 counts |
|
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) |
|
|
Current DC |
|
24.000 mA (Source) |
0.02%+ 2 counts |
|
24.000 mA (Simulate) |
0.02%+ 2 counts |
|
|
Resistance |
|
15.0 to 400.0 Ω |
0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
|
401 to 1500 Ω |
0.5 Ω(excitation current 0.05 to 0.8 mA) |
|
1500 to 3200 Ω |
1 Ω(excitation current 0.05 to 0.4 mA) |
|
|
Frequency |
|
2.0 to 1000.0 CPM |
0.05% |
|
1.0 to 1100.0 Hz |
0.05% |
|
1.00 to 10.00 kHz |
0.25% |
|
Waveform: |
5 V p-p squarewave, -0.1 V offset |
|
|
RTDs and Thermocouples |
|
Measure Accuracy: |
|
NI-120 |
0.2 °C |
|
PT-100 (385) |
0.33 °C |
|
PT-100 (393) |
0.3 °C |
|
PT-100 (JIS) |
0.3 °C |
|
PT-200 (385) |
0.2 °C |
|
PT-500 (385) |
0.3 °C |
|
PT-1000 (385) |
0.2 °C |
|
Resolution: |
0.1 °C |
|
J: |
0.7 °C |
|
K: |
0.8 °C |
|
T: |
0.8 °C |
|
E: |
0.7 °C |
|
R: |
1.8 °C |
|
S: |
1.5 °C |
|
B: |
1.4 °C |
|
L: |
0.7 °C |
|
U: |
0.75 °C |
|
N: |
0.9 °C |
|
Resolution |
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F |
|
XK |
0.6 °C |
|
BP |
1.2°C |
|
|
Source Accuracy: |
|
NI-120 |
0.2 °C |
|
PT-100 (385) |
0.33 °C |
|
PT-100 (393) |
0.3 °C |
|
PT-100 (JIS) |
0.3 °C |
|
PT-200 (385) |
0.2 °C |
|
PT-500 (385) |
0.3 °C |
|
PT-1000 (385) |
0.2 °C |
|
Resolution: |
0.1 °C |
|
Note: |
Accuracy stated for 4-wire measurement. |
|
J: |
0.7 °C |
|
K: |
0.8 °C |
|
T: |
0.8 °C |
|
E: |
0.7 °C |
|
R: |
1.4 °C |
|
S: |
1.5 °C |
|
B: |
1.4 °C |
|
L: |
0.7 °C |
|
U: |
0.75 °C |
|
N: |
0.9 °C |
|
Resolution |
J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C |
|
XK |
0.6 °C |
|
BP |
1.2°C |
|
|
Specifications |
|
Ramp Functions |
|
Source functions: |
Voltage, current, resistance, frequency, temperature |
|
Ramps |
Slow ramp, Fast ramp, 25% step-ramp |
|
|
Loop Power Function |
|
Voltage: |
24 V |
|
Accuracy: |
10% |
|
Maximum current: |
22 mA, short circuit protected |
|
|
Step Functions |
|
Source functions: |
Voltage, current, resistance, frequency, temperature |
|
Steps |
25% of range, 100% of range |
|
|
Environmental Specifications |
|
Operating Temperature |
|
|
Storage Temperature |
|
|
Operating Altitude |
|
|
Safety Specifications |
|
Agency Approvals |
|
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 |
|
|
Mechanical & General Specifications |
|
Size |
|
|
Weight |
|
|
Batteries |
|
|
Battery Replacement |
|
Separate battery compartment, accessable without breaking calibration seal |
|
|
Side port connections |
|
Pressure module connector, also used for remote real-time programming |
|