Specifications |
|
Vertical deflection |
|
Bandwidth |
100 MHz |
|
Real-time sample rate |
1.25 GS/s |
|
Channels |
2 |
|
Number of scope inputs |
2 input channels plus external trigger |
|
Rise Time |
3.5 |
|
Accuracy |
± (2.1 % of reading + 0.04 x range/div) @ 5 mV/div to 100 V/div |
|
Channel architecture |
All inputs fully insulated from each other and from ground Inputs may be activated in any combination |
|
Input coupling |
AC or DC, with ground level indicator |
|
Input sensitivity |
2 mV/div to 100 V/div, plus variable attenuation |
|
Bandwidth limiter |
User selectable: 20 kHz, 20 MHz or full bandwidth |
|
Normal/invert/variable |
On each input channel, switched separately |
|
Input voltage |
CAT III 1000 V/CAT IV 600 V rated, see General Specifications for further details |
|
Input impedance |
1 MΩ ± 1 % // 14 pF ± 2 pF |
|
Vertical resolution |
8 bit |
|
Horizontal deflection |
Max. real-time sample rate
(sampled simultaneously) |
1.25 GS/s for each channel |
|
Record length |
Up to 10,000 samples per channel |
|
Time base range |
5 ns/div to 4 s/div |
Time base in a 1-2-4-sequence
Slower time/division settings using ScopeRecord™ Roll mode |
|
Glitch capture |
8 ns |
Scope
measurements |
Cursor |
7 |
|
Automatic |
30 |
Maximum
Record length |
ScopeRecord mode |
30,000 pts per input |
|
Scope mode |
10,000 pts per input |
|
Timing accuracy |
± (0.01 % of reading + 1 pixel) |
|
Meter Modes |
|
Meter inputs |
Via 4 mm banana inputs, fully isolated from scope inputs and scope ground |
|
Number of readings |
One at a time |
|
Maximum resolution |
5,000 counts |
|
Input impedance |
1 MΩ ± 1 % // 14 pF ± 2 pF |
|
Advanced meter functions |
Auto/manual ranging, relative measurements (Zero reference), TrendPlot recording |
|
Horizontal deflection |
|
Vdc accuracy |
± (0.5 % + 5 counts) |
|
Vac true rms accuracy |
15 Hz to 60 Hz |
± (1 % + 10 counts) |
|
60 Hz to 1 kHz |
± (2.5 % + 15 counts) |
|
Vac+dc true rms accuracy |
15 Hz to 60 Hz |
± (1 % + 10 counts) |
|
60 Hz to 1 kHz |
± (2.5 % + 15 counts) |
|
Voltmeter ranges |
500 mV, 5 V, 50 V, 500 V, 1,000 V |
|
Resistance |
|
Ranges |
500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ |
|
Accuracy |
± (0.6 % + 5 counts) |
|
Other meter functions |
|
Continuity |
Beeper on < 50 Ω (± 30 Ω) |
|
Diode test |
Up to 2.8 V |
|
Current (A) |
Adc, Aac, Aac+dc using an optional current clamp or shunt Scaling factors: 0.1 mV/A, 1 mV/A to 100 V/A and 400 mV/A |
|
Temperature |
With optional accessories. Scale factors 1 °C/mV or 1 °F/mV |