รายละเอียดสินค้า Fluke 271 10 MHz DDS Function Generator
|
Waveforms |
|
Frequency |
|
|
All waveforms are available up to 10 MHz. However, the purity of triangle, ramp, and multi-level square wave waveforms is not specified above the frequencies indicated in the following section. |
|
Range |
0.1 mHz to 10 MHz |
|
Resolution |
7 digits or 0.1 mHz |
|
Accuracy |
Typically < ± 10 ppm for 1 year, 18 °C to 28 °C |
|
Tempco. |
Typically < 1 ppm/°C outside 18 °C to 28 °C |
|
|
Sinewave |
|
Distortion |
< –60 dBc to 20 kHz, < –50 dBc to 300 kHz, < -35 dBc to 10 MHz |
|
Spurii |
Non -harmonically related spurii typically <- 60 dBc to 10 MHz |
|
|
Squarewave |
|
Rise and fall times |
< 22 ns |
|
|
Triangle |
|
Linearity error |
< 0.5 % to 30 kHz |
|
|
Positive and Negative Ramp |
|
Linearity error |
< 0.5 % to 30 kHz |
|
|
Positive and Negative Pulse |
|
Rise and fall times |
<22 ns |
|
|
Multi-Level Squarewave |
|
|
Up to 16 steps available per cycle, each step selectable for amplitude (10 bit resolution) and duration (1 to 1024 samples). Above 27 kHz a 36 ns edge uncertainty is introduced. |
|
Rise and fall times |
< 22 ns |
|
|
Arbitrary (and complex) |
|
|
A number of “complex” waveforms are pre-programmed in ROM . A further five, user defined, waveforms may be loaded via the digital interfaces and stored in non-volatile RAM. Frequency range: All waveform points can be continuously output up to 27 kHz, beyond which they are sampled. |
|
No. of samples |
No. of samples |
|
Noise |
Wideband noise with variable amplitude and offset. |
|
|
Symmetry |
|
Range |
Sine — 1 % to 99 % at all frequencies; Other waveforms — 1 % to 99 % to 30 kHz,
20 % to 80 % to 10 MHz |
|
Resolution |
0.1 % |
|
|
Main Output |
|
Output impedance |
50 Ω or 600 Ω switchable |
|
Amplitude |
5 mV to 20 V pk-pk open circuit (2.5 mV to 10 V into 50 Ω/600 Ω). Output can be specified as V-H: 2
(open circuit value) or V (Voltage into the characteristic impedance) in pk-pk, RMS or dBm.
Note that in positive or negative pulse modes the amplitude range is 2.5 mV to 10 V pk-pk O/C. |
|
Accuracy |
Typically ±3 % ±1 mV at 1 kHz into 50 Ω/600 Ω |
|
Flatness |
±0.2 dB to 500 kHz; ±1 dB to 10 MHz |
|
Pulse aberrations |
<5 %+ 2 mV |
|
DC offset |
± 10 V from 50Ω/600 Ω offset plus signal peak limited to ± 10 V from 50 Ω/600 Ω |
|
Resolution |
3 digits or 1 mV for both amplitude and offset |
|
Modulation |
|
Amplitude Modulation |
|
Carrier frequency |
0.1 mHz to 10 MHz |
|
Carrier waveforms |
All |
|
Depth |
0 to 100 %, resolution 1 % |
|
Internal source |
1 kHz fixed sinewave or 0.005 Hz to 50 kHz square wave |
|
External |
See “VCA In” section |
|
|
Frequency Shift Keying (FSK) |
|
|
Phase coherent switching between two frequencies at a rate defined by the switching signal source |
|
Carrier frequency |
0.1 mHz to 10 MHz |
|
Carrier waveforms |
All |
|
Switch repetition rate |
dc to 50 kHz internal, dc to 1 MHz external |
|
Switching signal source. |
Internal from keyboard or trigger generator. External from EXT TRIG input or remote interface |
|
|
Operating Modes |
|
Trigger/burst |
|
|
Phase coherent signal keying — each positive edge of the trigger signal will produce one burst of the carrier, starting and stopping at the phase angle specified by the start/stop phase setting |
|
Carrier frequency |
0.1 mHz to 10 MHz |
|
Carrier waveforms |
All |
|
Number of cycles |
1 to 1023 (resolution 1 cycle) or 0.5 to 511.5 (resolution 1/2 cycle) |
|
Trigger rep. rate |
dc to 50 kHz internal, dc to 1 MHz external |
|
Source |
Internal from keyboard or trigger generator. External from EXT TRIG input or remote interface |
|
|
Gated |
|
|
Non phase-coherent signal keying — output is On while Gate signal is high and Off while low. |
|
Carrier frequency |
From 0.1 mHz to 10 MHz |
|
Carrier waveforms |
All |
|
Trigger rep. rate |
dc to 50 kHz internal dc to 1 MHz external |
|
Gate source |
Internal from keyboard or trigger generator. External from EXT TRIG input or remote interface |
|
|
Sweep |
|
Carrier waveforms |
All |
|
Sweep mode |
Linear or logarithmic, single or continuous |
|
Sweep width |
0.1 mHz to 10 MHz. Phase continuous. Independent setting of the start and stop frequency. |
|
Sweep time |
10 ms to 999 s (3 digit resolution) |
|
Markers |
Two markers variable during sweep. Available at the TRIG/SWEEP OUT socket |
|
Sweep trigger source |
The sweep may be free run or triggered from: keyboard, EXT TRIG input, remote interface |
|
|
Hop |
Up to 16 different “hop” waveforms can be defined in terms of function, frequency, amplitude, offset and duration. Duration setable per step 1 ms to 60 s. |
|
Start/Stop Phase |
|
Carrier frequency |
0.1 mHz to at least 1 MHz |
|
Carrier waveforms |
All |
|
Range |
–360 to +360 degrees |
|
Resolution |
1 degree |
|
Accuracy |
Typically 1 degree to 30 kHz |
|
|
Trigger Generator |
Internal source 0.005 Hz to 50 kHz squarewave adjustable in 20 us steps. 3 digit resolution. Available for external use from TRIG/SWEEP OUT socket. |
|
Auxiliary Outputs |
|
Aux Out |
CMOS/TTL levels with symmetry and frequency of main output and phase of start-stop phase setting |
|
Trig/Sweep Out |
Multi-function output depending upon mode. Except in sweep mode, the output is that of the trigger generator at CMOS/TTL levels from 1 kW.
In Sweep mode the output is a 3-level waveform, changing from high (+4 V) to low (0 V) at the start of sweep, with narrow 1 V pulses at each marker point. |
|
Inputs |
|
Ext Trig |
|
Frequency range |
DC to 1 MHz |
|
Signal range |
TTL (1.5 V) threshold; maximum input ± 10 V |
|
Min. pulse width |
50 ns |
|
|
VCA In |
|
Frequency range |
DC - 100 kHz |
|
Signal range |
2.5 V for 100 % level change at maximum output |
|
Input impedance |
Typically 6 kΩ |
|
|
Phase Locking |
|
Clock in/out |
TTL/CMOS threshold levels; output impedance typically 50 Ω as an output |
|
Sync out |
TTL/CMOS logic levels from typically 50 Ω. The signals from these sockets are used to phase lock two or more generators. |
|
Interfaces |
|
RS-232 |
Variable Baud rate, 9600 Baud maximum. 9-pin D-connector. |
|
IEEE-488 |
Conforming with IEEE488.1 and IEEE488.2 |
|
General Specifications |
|
Display |
20 character x 4 row alphanumeric LCD |
|
Data entry |
Keyboard selection of mode, waveform etc.; value entry direct by numeric keys or by rotary control. |
|
Stored settings |
Up to 9 complete instrument set-ups may be stored and recalled from battery-backed memory. |
|
Size |
3U (130 mm) height; half-rack (212 mm) width, 330 mm long |
|
Weight |
4.1 kg (9 lb) |
|
Power |
100 V ac, 110 to 120 V ac or 220 to 240 V ac +/– 10 %, 50/60 Hz ac by internal adjustment; 30 VA max. |
|
Operating range |
+5 °C to 40 °C, 20 to 80 % RH |
|
Storage range |
–20 °C to +60 °C |
|
Options |
IEEE-488 interface; 19-in rack mounting kit |